Calculations of the resistance values of 20 thermistors at 0°C

Description

The presented data set is part of the research aimed at determining the actual characteristics of each thermistor in a package of twenty NTC10k type sensors.

The measurement system is shown in Fig. 1. A measurement system based on NI PXIe-1082Q (National Instruments Chassis), equipped with NI PXIe-8135 (Embedded Controller), NI PXIe-6341 (DAQ card) and NI SCB-68A (68-pin connector block), was used for the acquisition of measurement data. Eleven input channels of the DAQ card were used in the measurements. Channel 0 (ch_0) was used to measure the supply voltage. Sensors numbered 1 to 20 were tested in two parts, ten sensors per set. They were connected to the inputs of channels 1-10 (ch_1-ch_10) of the DAQ card.

The conditioning unit for individual sensors is shown in Fig.2. These circuits are performed on connector block SCB-68A as the voltage dividers. For the set of ten thermistors, the supply voltage is measured in channel 0 (ch_0) of the DAQ card, and individual thermistors are connected to channels 1-10 (ch_i). 

The thermistor set was placed in a thermal tank containing a mixture of water and ice at a temperature of 0°C (confirmed by a digital temperature gauge DT-34 with an accuracy of 0.1°C).

The calculations of the resistance values were performed based on the datasets [ Maśnicki, R., Tyndzik, K., & Pałczyńska, B. (2021). Calibration of NTC 10k thermistors in temperature of 0 C [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/dcn8-z120].

Illustration of the publication

Fig. 1. The measurement setup

The mentioned datasets consist of 15  files 0.0Cx.lvm  and 15 files II0.0Cx.lvm (where: x is from 0 to 14), containing the measurement results supply voltages Uz (ch_0) (Fig. 2) and voltages at the outputs of ten voltage dividers (ch_1 to ch_10), repeated in two series (at 0.0°C, for two batches of thermistors, 10 per batch) with 30 measurements for each of the 11 channels. The resistance calculations were performed for 20 thermistors. The file Rt0.0C.lvm contains data for the first part (ten thermistors), and file RtII0.0C.lvm contains data for the second part (another ten thermistors) of 20 thermistors set. The last two lines show the averaged resistance values with standard deviations.

Illustration of the publication

                                                                                                              Fig. 2. The sensor conditioning unit

Dataset file

Resistance_0.0C.zip
2.9 kB, S3 ETag 93223044711ad367437481bee4676909-1, downloads: 65
The file hash is calculated from the formula
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} where a single part of the file is 512 MB in size.

Example script for calculation:
https://github.com/antespi/s3md5
download file Resistance_0.0C.zip

File details

License:
Creative Commons: by-nc 4.0 open in new tab
CC BY-NC
Non-commercial

Details

Year of publication:
2021
Verification date:
2021-05-31
Creation date:
2021
Dataset language:
English
Fields of science:
  • Automation, electronic and electrical engineering (Engineering and Technology)
DOI:
DOI ID 10.34808/2ztn-9f40 open in new tab
Series:
Verified by:
Gdańsk University of Technology

Keywords

References

Cite as

seen 137 times