Electrochemical measurements of FLBP modifications using isocyanates - Open Research Data - Bridge of Knowledge

Search

Electrochemical measurements of FLBP modifications using isocyanates

Description

The electrochemical behavior of modified FLBP was investigated using a potentiostat-galvanostat (VMP-300, Bio-Logic, France) with EC-Lab software (v11.50). Measurements were carried out in a three-electrode configuration (Pt wire as the counter electrode, Ag/AgCl wire as a pseudo-reference electrode) under an inert atmosphere. 

 A 5 mM solution of the inner-sphere redox mediator [Fe(CN)₆]³⁻/⁴⁻ in 1 M KCl was used. This solution was prepared by mixing two salts: 2.5 mM K₃[Fe(CN)₆] and 2.5 mM K₄[Fe(CN)₆].

Cyclic voltammetry (CV) tests were conducted within a potential range of -0.25 V to +0.7 V (vs Ag/AgCl in 1 M KCl) at various scan rates: 5 mV s⁻¹, 10 mV s⁻¹, 25 mV s⁻¹, 50 mV s⁻¹, 100 mV s⁻¹, 150 mV s⁻¹, 200 mV s⁻¹, and 300 mV s⁻¹.

Electrochemical impedance spectroscopy (EIS) was performed by measuring the system’s electrical response at its open-circuit potential. The measurements covered a broad frequency range from 0.01 Hz to 100 kHz, with a sinusoidal amplitude of 10 mV.

Dataset file

electrochemical measurements.zip
65.7 MB, S3 ETag f817dd3fd68948ea7c9b3a543ed46846-1, downloads: 0
The file hash is calculated from the formula
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} where a single part of the file is 512 MB in size.

Example script for calculation:
https://github.com/antespi/s3md5
download file electrochemical measurements.zip

File details

License:
Creative Commons: by 4.0 open in new tab
CC BY
Attribution
Raw data:
Data contained in dataset was not processed.
Software:
EC-Lab

Details

Year of publication:
2025
Verification date:
2025-05-16
Creation date:
2023
Dataset language:
English
DOI:
DOI ID 10.34808/98a9-vq18 open in new tab