Testing HIADAC high impedance analyzer in cooperation with Road and Bridge Research Institute on polyvinyl coating using 2-wire probe sample 1
Description
The dataset presents impedance spectrum of polyvinyl coating (code-name PW_2_2_1) sample 1. This object was used to test high-impedance analyzer for diagnostic of anticorossion coatings (HIADAC) realized in the frame of Eureka project E!3174. The impedance spectrum frequency range (0.1 Hz – 100 kHz) was selected in order to test the whole measureement range of the analyzer. In this dataset, the DC polarisation was to 0V and AC amplitude was set to 1 V. The model together with the 2-wire measurement probe (dedicated for grounded objects) was placed in the shielding box acting as Faraday cage. The cells, measurement setup and shielding box are presented in the photos below.
Top-left figure presents measurement setup (on the left HIADAC is located), top-right figure presents shielding box for the measurements (with cell and measurement probe in side) and bottom-left figure presents cell mounted on the coatings under the test.
Dataset file
hexmd5(md5(part1)+md5(part2)+...)-{parts_count}
where a single part of the file is 512 MB in size.Example script for calculation:
https://github.com/antespi/s3md5
File details
- License:
-
open in new tabCC BY-NCNon-commercial
Details
- Year of publication:
- 2021
- Verification date:
- 2021-04-30
- Dataset language:
- English
- Fields of science:
-
- Automation, electronic and electrical engineering (Engineering and Technology)
- DOI:
- DOI ID 10.34808/91n2-mr19 open in new tab
- Series:
- Verified by:
- Gdańsk University of Technology
Keywords
Cite as
Authors
seen 107 times