Effect of oxygenation time on signal of a sensor based on ionic liquids - Publikacja - MOST Wiedzy

Wyszukiwarka

Effect of oxygenation time on signal of a sensor based on ionic liquids

Abstrakt

The paper presents an oxygen sensor based on ionic liquids and solid electrodes. The following ionic liquids have been employed: [BMIM][BF4], [HMIM][Cl], [BMIM][N(CN2)]. Minimum time of the sensor exposure to analyte, after which the signal (current intensity) was stable, has been evaluated. An impact of volumetric flow rate of analyte on the sensor exposure time and signal has been determined. A product of permeability coefficient and solubility of oxygen in ionic liquids has been estimated. A mechanism of oxygen reduction on a surface of the solid electrodes, in the ionic liquid environment has been presented. Overall cathodic coefficients of transition for the sensors with particular ionic liquids have been determined as a function of potential scan rate.

Cytuj jako

Pełna treść

pełna treść publikacji nie jest dostępna w portalu

Słowa kluczowe

Informacje szczegółowe

Kategoria:
Publikacja w czasopiśmie
Typ:
artykuł w czasopiśmie wyróżnionym w JCR
Opublikowano w:
ELECTROCHIMICA ACTA nr 56, strony 9910 - 9915,
ISSN: 0013-4686
Język:
angielski
Rok wydania:
2011
Opis bibliograficzny:
Gębicki J., Kloskowski A., Chrzanowski W.: Effect of oxygenation time on signal of a sensor based on ionic liquids// ELECTROCHIMICA ACTA. -Vol. 56, nr. iss. 27 (2011), s.9910-9915
Weryfikacja:
Politechnika Gdańska

wyświetlono 145 razy

Publikacje, które mogą cię zainteresować