Abstract
Example S13 from EA-4/02 M:2013 “Calibration of a ring gauge with a nominal diameter of 90 mm” is analyzed. The report uses the kurtosis method and the law of propagation of expanded uncertainty developed by the authors to the expanded uncertainty evaluation. It is shown that the introduction of a coaxiality correction for the ring gauge and the measuring axis of the comparator leads to the need to estimate its standard measurement uncertainty using the second-order terms of the Taylor series and taking into account the kurtosis of input quantities using the method of partial increments. A good agreement between the results obtained by applying the described procedure and the results obtained by the Monte Carlo method is shown.
Authors (4)
Cite as
Full text
full text is not available in portal
Keywords
Details
- Category:
- Conference activity
- Type:
- publikacja w wydawnictwie zbiorowym recenzowanym (także w materiałach konferencyjnych)
- Language:
- English
- Publication year:
- 2023
- Bibliographic description:
- Zakharov I., Botsiura O., Diakov D., Świsulski D.: Peculiarities of measurement uncertainty evaluation at calibrating a ring gauge// / : , 2023,
- Sources of funding:
-
- Free publication
- Verified by:
- Gdańsk University of Technology
seen 78 times
Recommended for you
Accounting for the distributions of input quantities in the procedure for the measurement uncertainty evaluation when calibrating the goniometer
- I. Zakharov,
- I. Zadorozhna,
- D. Świsulski
- + 1 authors
Uncertainty of mass flow measurement using centric and eccentric orifice for Reynolds number in the range 10,000 ≤ Re ≤ 20,000
- A. Golijanek-Jędrzejczyk,
- A. Mrowiec,
- R. Hanus
- + 2 authors