Abstract
In this paper, a low-coherence Fabry-Perot sensor with a spectrally measured signal processing response to the refractive index of liquids is presented. Optical fiber sensors are potentially capable of continuous measuring hematocrit levels in blood. Low-coherence Fabry-Perot interferometric sensors offer a robust solution, where information about the measurand is encoded in the full spectrum of light reflected from the sensing interferometer. The first step in the research on such sensor is the assessment of its performance under favorable conditions, i.e., using blood samples from healthy volunteers tested in vitro. Such an experiment was conducted using a sensor comprising a superluminescent diode source, an optical spectrum analyzer working as the detection setup and a sensing Fabry-Perot interferometer providing high interference contrast. The response of this sensor was recorded for several samples and compared with the reference laboratory method. The coefficient of determination (R2) for a linear relationship between the results given by both methods was 0.978 and the difference between these results was less than 1%. The presented results suggest that further research into the performance of the sensor is merited.
Author (1)
Cite as
Full text
- Publication version
- Accepted or Published Version
- License
- open in new tab
Keywords
Details
- Category:
- Articles
- Type:
- artykuł w czasopiśmie wyróżnionym w JCR
- Published in:
-
SENSORS
no. 14,
edition 4,
pages 6965 - 6976,
ISSN: 1424-8220 - Language:
- English
- Publication year:
- 2014
- Bibliographic description:
- Jędrzejewska-Szczerska M.: Response of a New Low-Coherence Fabry-Perot Sensor to Hematocrit Levels in Human Blood// SENSORS-BASEL. -Vol. 14, iss. 4 (2014), s.6965-6976
- Verified by:
- Gdańsk University of Technology
Referenced datasets
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 0 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 40 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 30 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 200 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 80 um (serie 1)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 170 um (serie 1)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 20 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 140 um (serie 1)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 130 um (serie 2)
- dataset Distance measurement with the low coherent interferometer with silver mirror (the source wavelegth 1310 nm) - 120 um (serie 2)
seen 141 times