Displacements of bones during bending test of first metatarsophalangeal joint after arthrodesis with medially or dorsally positioned locking plate and lag screw.

Description

The Dataset contains the values of displacements of bone control points  during the bending test of first metatarsophalangeal (MTP1) joint specimens after arthrodesis.

Three groups of samples were created depending on the MTP1 arthrodesis method:

-   group with medially positioned locking plate, samples P01-P02, P04-P07, P22;
-   group with dorsally positioned locking plate and lag screw, samples P08-P14;
-   group with medially positioned locking plate and lag screw, samples P15-P21.

T
he same short titanium locking plate (Medgal) and six 2.4 mm locking screws were used in all groups. The artificial models of metatarsal bone and proximal phalanx were created based on freely accessible STL geometry of the human foot (https://www.thingiverse.com/thing:22628). The geometry was scaled and cut to prepare bones for arthrodesis and fixation in dedicated steel jigs. The artificial bone models were additively manufactured with the fused filament fabrication (FFF) method using polylactide (PLA) material in Ultimaker 3 Extended 3D printer.
Illustration of the publication

The bending test of total 21 samples was performed in a Zwick/Roell Z10 universal testing machine (UTM).  The shaft of the metatarsal bone was clamped in stainless steel jig, while the“distal” end of the proximal phalanx was fixed in stainless steel clamp, which was connected to a traverse of UTM. A preload of 1 N and constant vertical displacement increment of 5 mm/min of the traverse were applied in the tests to simulate quasi static loading conditions. The tests were stopped after specimens failure. Digital image correlation (DIC) method and ARAMIS MC 3D 12M system were used to measure displacements at three control points P1, P2 and P3, see Figure.  The displacements between points P1 and P2 allow to calculate the relative displacements between the bones.

The  file has (.txt) extension and contains: 

- the time of the test (in s, 1nd column);
- displacement in X direction of point P1 (in mm, 2nd column), P2 (in mm, 3rd column), P3 (in mm, 4rd column);
- displacement in Y direction of point P1 (in mm, 5th column), P2 (in mm, 6th column), P3 (in mm, 7th column);
- displacement in Z direction of point P1 (in mm, 8th column), P2 (in mm, 9th column), P3 (in mm, 10th column);
- force (in N, 11th column).

Dataset file

Displacements_bones_bending.zip
181.7 kB, S3 ETag be324cfc5671714559896929cd1918d7-1, downloads: 5
The file hash is calculated from the formula
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} where a single part of the file is 512 MB in size.

Example script for calculation:
https://github.com/antespi/s3md5
download file Displacements_bones_bending.zip

File details

License:
Creative Commons: by 4.0 open in new tab
CC BY
Attribution
File embargo:
2023-12-31
Raw data:
Data contained in dataset was not processed.

Details

Year of publication:
2022
Verification date:
2022-11-17
Dataset language:
English
Fields of science:
  • Civil engineering and transport (Engineering and Technology)
  • Biomedical engineering (Engineering and Technology)
DOI:
DOI ID 10.34808/wqxn-8z65 open in new tab