Maximum forces in tendon connections for an anchor without pin in the single-side sample experiment - Open Research Data - Bridge of Knowledge

Search

Maximum forces in tendon connections for an anchor without pin in the single-side sample experiment

Description

The main objective of the all study was to assess the load capacity of a new implant device for knotless tendon anastomosis. During this series of experiments the load capacity of the anchoring element without a pin (see figure 1) was examined. Tendons from fresh swine legs (commercially available) served as the biological material. All samples were subjected to a load-to-failure test in the single-side sample experiment (see figure 2). They were axially stretched with a constant elongation increase of 0.5 mm/s. The tests were conducted on the custom-designed TissueTester 1.0 (see figure 3).

Illustration of the publication
Illustration of the publication
Illustration of the publication

Dataset file

DATA1.ZIP
242.6 kB, S3 ETag ce29bf7c396dd4b98a83f31820cbffb4-1, downloads: 61
The file hash is calculated from the formula
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} where a single part of the file is 512 MB in size.

Example script for calculation:
https://github.com/antespi/s3md5
download file DATA1.ZIP

File details

License:
Creative Commons: by-nc-nd 4.0 open in new tab
CC BY-NC-ND
Non-commercial - No Derivative Works

Details

Year of publication:
2021
Verification date:
2021-05-27
Dataset language:
English
Fields of science:
  • mechanical engineering (Engineering and Technology)
  • biomedical engineering (Engineering and Technology)
  • medical sciences (Medical and Health Sciences )
DOI:
DOI ID 10.34808/xvre-6372 open in new tab
Series:
Verified by:
Gdańsk University of Technology

Keywords

Cite as

seen 126 times