Analisys of the failure of fixator used in bone surgery - Publication - Bridge of Knowledge

Search

Analisys of the failure of fixator used in bone surgery

Abstract

The Gamma nail (fixator) made of stainless steel, which has broken in the body, was subjected to examination. The fixator was removed from the patient femur. The fracture occurred in the area of the nail hole after two months' stay in the organism. The research included chemical composition, hardness of the Gamma nail material and the observation and analysis of the type of fracture. The research showed the fatigue failure of the Gamma nail

Cite as

Full text

download paper
downloaded 24 times
Publication version
Accepted or Published Version
License
Creative Commons: CC-BY-NC-ND open in new tab

Keywords

Details

Category:
Articles
Type:
artykuły w czasopismach recenzowanych i innych wydawnictwach ciągłych
Published in:
Advances in Materials Science no. 8, pages 84 - 88,
ISSN: 1730-2439
Language:
English
Publication year:
2008
Bibliographic description:
Świeczko-Żurek B., Serbiński W., Szumlański A.: Analisys of the failure of fixator used in bone surgery// Advances in Materials Science. -Vol. 8., nr. nr 2=16 (2008), s.84-88
Verified by:
Gdańsk University of Technology

seen 95 times

Recommended for you

Meta Tags