Abstract
The Gamma nail (fixator) made of stainless steel, which has broken in the body, was subjected to examination. The fixator was removed from the patient femur. The fracture occurred in the area of the nail hole after two months' stay in the organism. The research included chemical composition, hardness of the Gamma nail material and the observation and analysis of the type of fracture. The research showed the fatigue failure of the Gamma nail
Authors (3)
Cite as
Full text
download paper
downloaded 24 times
- Publication version
- Accepted or Published Version
- License
- open in new tab
Keywords
Details
- Category:
- Articles
- Type:
- artykuły w czasopismach recenzowanych i innych wydawnictwach ciągłych
- Published in:
-
Advances in Materials Science
no. 8,
pages 84 - 88,
ISSN: 1730-2439 - Language:
- English
- Publication year:
- 2008
- Bibliographic description:
- Świeczko-Żurek B., Serbiński W., Szumlański A.: Analisys of the failure of fixator used in bone surgery// Advances in Materials Science. -Vol. 8., nr. nr 2=16 (2008), s.84-88
- Verified by:
- Gdańsk University of Technology
seen 95 times
Recommended for you
Fracture surface topography investigation and fatigue life assessment of notched austenitic steel specimens
- W. Macek,
- G. Robak,
- K. Żak
- + 1 authors
2022
A brief note on monotonic and fatigue fracture events investigation of thin-walled tubular austenitic steel specimens via fracture surface topography analysis (FRASTA)
- W. Macek,
- D. Sampath,
- Ł. Pejkowski
- + 1 authors
2022
Effect of bending-torsion on fracture and fatigue life for 18Ni300 steel specimens produced by SLM
- W. Macek,
- R. Masoudi Nejad,
- S. Zhu
- + 5 authors
2023