Investigation of poly(3,4-ethylenedioxythiophene) deposition method influence on properties of ion-selective electrodes based on bis(benzo-15-crown-5) derivatives
Abstract
Glassy carbon electrodes modified by conductive polymers and membrane with derivatives of bis(benzo-15-crown-5) were tested as solid contact ion selective electrodes for K+ ions concentration determination. PEDOT with PSS, Cl- and ClO4- counter ions was electrochemically deposited onto glassy carbon substrates using four different electrochemical approaches (potentiostatic, galvanostatic, potentiodynamic and potentiostatic pulses). Scanning electron microscopy was applied to investigate influence of electrodeposition method on morphology of polymer films and drop-casted pseudo liquid membrane with an ionophore. The presence of the polymer film and method of deposition affect morphology of tested electrodes and sensing properties as well. The best sensing properties were obtained for the electrodes with biscrown as ionophore with polymer which have a developed surface prepared via potentiostatic pulses (logKK/Na = -5.6 using separate solution method (SSM)). Galvanostatic method of electrodeposition which resulted in relatively smooth layer of the PEDOT exhibited the poorest sensing properties for sensor with biscrown as ionophore (logKK/Na = -4.3 for SSM). All prepared sensors exhibited very favorable values of selectivity coefficients KK/Na (log KK/Na= -5.6 - -4.0). Detection limit equals to 10-7 M was achieved. The influence of pH for sensing properties were investigated for the selected electrodes.
Citations
-
7
CrossRef
-
0
Web of Science
-
8
Scopus
Authors (4)
Cite as
Full text
- Publication version
- Accepted or Published Version
- License
- open in new tab
Keywords
Details
- Category:
- Articles
- Type:
- artykuł w czasopiśmie wyróżnionym w JCR
- Published in:
-
ELECTROCHIMICA ACTA
no. 246,
pages 424 - 432,
ISSN: 0013-4686 - Language:
- English
- Publication year:
- 2017
- Bibliographic description:
- Jeszke M., Trzciński K., Karczewski J., Luboch E.: Investigation of poly(3,4-ethylenedioxythiophene) deposition method influence on properties of ion-selective electrodes based on bis(benzo-15-crown-5) derivatives// ELECTROCHIMICA ACTA. -Vol. 246, (2017), s.424-432
- DOI:
- Digital Object Identifier (open in new tab) 10.1016/j.electacta.2017.05.197
- Bibliography: test
-
- U.E. Spichiger-Keller, Chemical sensors and biosensors for medical and biological applications, Wiley-VCH, 1998. open in new tab
- K. Kimura, T. Maeda, H. Tamura, T. Shono, Potassium-Selective PVC Membrane Electrodes Based on Bis-and Poly(Crown Ether)s, J. Electroanal. Chem. 95 (1979) 91-101. open in new tab
- G.J. Moody, B.B. Saad, J.D.R. Thomas, Studies on Bis(crown ether)-based Ion- selective Electrodes for the Potentiometric Determination of Sodium and Potassium in Serum, Analyst 114 (1989) 15-20. open in new tab
- Z. Xia, I.H.A. Badr, S.L. Plummer, Lawrence Cullen, Leonidas G. Bachas, Synthesis and Evaluation of a Bis(crown ether) Ionophore with a Conformationally Constrained Bridge in Ion-Selective Electrodes, Analytical Sciences 14 (1998) 169-173. open in new tab
- J.L. Atwood, J.E. Davies, D.D. Macnicol, F. Vögtle, Comprehensive Supramolecular Chemistry, Molecular Recognition: Receptors for Cationic Guests, vol. 1(1996) , pp. p.605.
- M.A. Bush, M.R. Truter, Crystal structures of complexes between alkali-metal salts and cyclic polyethers. Part III. Aquo-(2,3-benzo-1,4,7,10,13- pentaoxacyclopentadec-2-ene)sodium iodide, J. Chem. Soc., Perkin Trans. 2 (1972) 341-344. open in new tab
- P.R. Mallison, M.R. Truter, Crystal structures of complexes between alkali- metal salts and cyclic polyethers. Part V. The 1: 2 complex formed between potassium iodide and 2,3,5,6,8,9,11,12-octahydro-1,4,7,10,13- benzopentaoxacyclopentadecin (benzo-15-crown-5), J. Chem. Soc., Perkin Trans. 2 (1972) 1818-1823. open in new tab
- M. Bourgoin, K.H. Wong, J.Y. Hui, J. Smid, Interactions of macrobicyclic polyethers with ions and ion pairs of picrate salts, J. Am. Chem. Soc. 97 (1975) 3462-3467. open in new tab
- E. Kleinpeter, I. Starke, D. Ströhl, H.J. Holdt, NMR Spectroscopic Study of Solution Structure and Complexational Behaviour of bis-benzo crown ethers, Journal of Molecular Structure 404 (1997) 273-290. open in new tab
- S. Fery-Forgues, F. Al-Ali, Bis(azacrown ether) and bis(benzocrown ether) Dyes: Butterflies Tweezers And Rods In Cation Binding, J. Photochem. Photobiol. C: Photochemistry Reviews 5 (2004) 139-153. open in new tab
- M. Szarmach, E. Wagner-Wysiecka, M.S. Fonari, E. Luboch, Bis(azobenzocrown ether)s -Synthesis and Ionophoric Properties, Tetrahedron 68 (2012) 507-515. open in new tab
- Y. Inoue, T. Hakushi, Y. Liu, L.-H. Tong, J. Hu, G.-D. Zhao, S. Huang, B.-Z. Tian, Complexation Thermodynamics of Bis(crown ether)s. Calorimetric Titration of Complexation of Potassium Ion with Bis(benzocrown ether)s: Enthalpy- Entropy Compensation, J. Phys. Chem. 92 (1988) 2371-2374. open in new tab
- E. Luboch, A. Cygan, J.F. Biernat, Bis(benzocrown ether)s with Polymethylene Bridges and Their Application in Ion-Selective Electrodes, Tetrahedron 47 (1991) 4101-4112. open in new tab
- E. Luboch, A. Cygan, J.F. Biernat, A. Yu. Simonov, A.A. Dvorkin, XVII International Symposium on Macrocyclic Chemistry, (1992) Provo, Utah, USA P10.
- E. Pungor, Ion-Selective Electrodes. 4. (1984) 82, U. Oesch, P., Anker, D. Ammann and W. Simon, Membrane Technological Optimization of Ion- Selective Electrodes Based on Solvent Polymeric Membranes for Clinical Applications, 4th Symposium of Ion-Selective Electrodes, Mátrafüred, 1984.
- J. Bobacka, A. Ivaska, A. Lewenstam, Potentiometric Ion Sensors Based on Conducting Polymers, Electroanalysis 15 (2003) 366-374. open in new tab
- A. Cadogan, Z. Gao, A. Lewenstam, A. Ivaska, D. Diamond, All-solid-state Sodium-Selective Electrode Based on a Calixarene Ionophore in a Poly(Vinyl Chloride) Membrane With a Polypyrrole Solid Contact, Anal. Chem. 64 (1992) 2496-2501. open in new tab
- E. Bakker, P. Buhlmann, E. Pretsch, Carrier-Based Ion-Selective Electrodes and Bulk Optodes. 1. General Characteristics, Chem. Rev. 97 (1997) 3083-3132. open in new tab
- P. Buhlmann, E. Pretsch, E. Bakker, Carrier-Based Ion-Selective Electrodes and Bulk Optodes. 2. Ionophores for Potentiometric and Optical Sensors, Chem. Rev. 98 (1998) 1593-1688. open in new tab
- F. Garnier, Functionalized Conducting Polymers-Towards Intelligent Materials, Angew. Chem. 101 (1989) 529-533. open in new tab
- J. Bobacka, T. Lindfors, M. McCarrick, A. Ivaska, A. Lewenstam, Single-Piece all- Solid-State Ion-Selective Electrode, Anal. Chem. 67 (1995) 3819-3823. open in new tab
- L.Y. Heng, E.A.H. Hall, One-Step Synthesis of K + -Selective Methacrylic-Acrylic Copolymers Containing Grafted Ionophore and Requiring No Plasticizer, Electroanalysis 12 (2000) 178-186. open in new tab
- J. Bobacka, Conducting Polymer-Based Solid-State Ion-Selective Electrodes, Electroanalysis 18 (2006) 7-18. open in new tab
- A. Michalska, K. Maksymiuk, Conducting Polymer Membranes for Low Activity Potentiometric Ion Sensing, Talanta 63 (2004) 109-117. open in new tab
- M. Vazquez, J. Bobacka, A. Ivaska, A. Lewenstam, Influence of Oxygen and Carbon Dioxide on the Electrochemical Stability of Poly(3,4- Ethylenedioxythiophene) Used as Ion-to-Electron Transducer in all-Solid- State Ion-Selective Electrodes, Sens. Actuators B 82 (2002) 7-13. open in new tab
- Z. Mousavi, J. Bobacka, A. Lewenstam, A. Ivaska, Poly(3,4- ethylenedioxythiophene) (PEDOT) Doped with Carbon Nanotubes as Ion-to- Electron Transducer in Polymer Membrane-Based Potassium Ion-Selective Electrodes, Journal of Electroanalytical Chemistry 633 (2009) 246-252. open in new tab
- M. Mir, R. Lugo, I.B. Tahirbegi, J. Samitier, Miniaturizable Ion-Selective Arrays Based on Highly Stable Polymer Membranes for Biomedical Applications, Sensors 14 (2014) 11844-11854. open in new tab
- M. Guzi nski, G. Lisak, T. Sokalski, J. Bobacka, A. Ivaska, M. Boche nska, A. Lewenstam, Solid-Contact Ion-Selective Electrodes with Highly Selective Thioamide Derivatives of p-tert-Butylcalix[4] arene for the Determination of Lead(II) in Environmental Samples, Anal. Chem. 85 (2013) 1555-1561. open in new tab
- M. Guzinski, J.M. Jarvis, B.D. Pendley, E. Lindner, Equilibration Time of Solid Contact Ion-Selective Electrodes, Anal. Chem. 87 (2015) 6654-6659. open in new tab
- Z. Mousavi, J. Bobacka, A. Ivaska, Potentiometric Ag + Sensors Based on Conducting Polymers: A Comparison between Poly(3,4- ethylenedioxythiophene) and Polypyrrole Doped with Sulfonated Calixarenes, Electroanalysis 17 (2005) 1609-1615. open in new tab
- M. Vazquez, J. Bobacka, M. Luostarinen, K. Rissanen, A. Lewenstam, A. Ivaska, Potentiometric Sensors Based on Poly(3,4-Ethylenedioxy-Thiophene) (PEDOT) Doped with Sulfonated calix[4] arene and calix[4] resorcarenes, J. Solid State Electrochem. 9 (2005) 312-319. open in new tab
- A. Michalska, A. Gałuszkiewicz, M. Ogonowska, M. Ocypa, K. Maksymiuk, PEDOT films: multifunctional membranes for electrochemical ion sensing, J. Solid State Electrochem. 8 (2004) 381-389. open in new tab
- E. Lindner, Y. Umezawa, Performance Evaluation Criteria For Preparation and Measurement of Macro-and Microfabricated Ion-Selective Electrodes, Pure Appl. Chem. 80 (2008) 85-104. open in new tab
- P.C. Meier, Two-Parameter Debye-Hueckel Approximati on for The Evaluation of Mean Activity Coefficients of 109 Electrolytes, Anal. Chim. Acta 136 (1982) 363-368. open in new tab
- B.E.H. Saxberg, R. Kowalski, Generalized Standard Addition Method, Analytical Chemistry 51 (1979) 1031-1038. open in new tab
- K. Trzci nski, M. Szkoda, K. Siuzdak, M. Sawczak, A. Lisowska-Oleksiak, Enhanced Photoelectrochemical Performance of Inorganic-Organic Hybrid Consisting of Bivo 4 and PEDOT:PSS, Appl. Surf. Sci. 388 (2016) 753-761. open in new tab
- J.A. Cobos-Murcia, L. Galicia, A. Rojas-Hernández, M.T. Ramírez-Silva, R. Álvarez-Bustamante, M. Romero-Romo, G. Rosquete-Pina, M. Palomar- Pardave, Electrochemical Polymerisation of 5-Amino-1, 10-Phenanthroline onto Different Substrates, Experimental and Theoretical Study, Polymer (Guildf). 46 (2005) 9053. open in new tab
- E. Garfias-García, M. Romero-Romo, M.T. Ramírez-Silva, J. Morales, M. Palomar-Pardavé, Eletrochemical Nucleation of Polypyrrole onto Different Substrates, Int. J. Electrochem. Sci. 5 (2010) 763-773. open in new tab
- J. Bobacka, Potential Stability of All-Solid-State Ion-Selective Electrodes Using Conducting Polymers as Ion-to-Electron Transducers, Anal. Chem. 71 (1999) 4932-4937. open in new tab
- G.J. Moody, Bahruddin B. Saad, J.D.R. Thomas, Studiesn Bis(Crown Ether)-Based Ion-Selective Electrodes for The Potentiometric Determination of Sodium and Potassium in Serum, Analyst 114 (1989) 15-20. open in new tab
- S. Chandra, R. Buschbeck, H. Lang, Triethylene Glycol Ether End-grafted Carbosilane Dendrimer: A Potential Ionophore for Potassium Ion Recognition, Analytical Sciences 22 (2006) 1327-1332. open in new tab
- D. Siswanta, Y.D. Wulandari, J. Jumina, Synthesis of Poly(Benzyl-eugenol) and Its Application as an Ionophore for a Potassium Ion-Selective Electrode, Eurasian Journal of Analytical Chemistry 11 (2016) 115-125.
- S.S.M. Hassan, W.H. Mahmoud, A.H.M. Othman, A Novel Potassium Ion Membrane Sensor Based on Rifamycin Neutral Ionophore, Talanta 44 (1997) 1087-1094. open in new tab
- W.S. Han, Y.H. Lee, K.J. Jung, S.Y. Ly, T.K. Hong, M.H. Kim, Potassium Ion- Selective Polyaniline Solid-Contact Electrodes Based on 4',4(5)-Di-tert butyldibenzo-18-crown-6-ether Ionophore, Journal of Analytical Chemistry 63 (2008) 987-993. open in new tab
- S. Kim, H. Kim, K.H. Noh, S.H. Lee, S.K. Kim, J.S. Kim, Potassium Ion-Selective Membrane Electrodes Based on 1,3-alternate calix[4] crown-5-azacrown-5, Talanta 61 (2003) 709-716. open in new tab
- H.J. Adrouge, N.E. Madias, Am. J. Med. 71 (1981) 456.
- Verified by:
- Gdańsk University of Technology
seen 119 times