Calculations of material factors for a set of 20 thermistors - Open Research Data - Bridge of Knowledge

Search

Calculations of material factors for a set of 20 thermistors

Description

The presented data set is part of the research aimed at determining the actual characteristics of each thermistor in a package of twenty NTC10k type sensors.

The measurement system is shown in Fig. 1. A measurement system based on NI PXIe-1082Q (National Instruments Chassis), equipped with NI PXIe-8135 (Embedded Controller), NI PXIe-6341 (DAQ card) and NI SCB-68A (68-pin connector block), was used for the acquisition of measurement data. Eleven input channels of the DAQ card were used in the measurements. Channel 0 (ch_0) was used to measure the supply voltage. Sensors numbered 1 to 20 were tested in two parts, ten sensors per set. They were connected to the inputs of channels 1-10 (ch_1-ch_10) of the DAQ card.

The conditioning unit for individual sensors is shown in Fig.2. These circuits are performed on connector block SCB-68A as the voltage dividers. For the set of ten thermistors, the supply voltage is measured in channel 0 (ch_0) of the DAQ card, and individual thermistors are connected to channels 1-10 (ch_i).

The calculations of the material factors were performed based on the following datasets [Maśnicki, R., Tyndzik, K., & Pałczyńska, B. (2021). Calibration of NTC 10k thermistors in temperature of 0 C [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/dcn8-z120], [Maśnicki, R., Tyndzik, K., & Pałczyńska, B. (2021). Calibration of NTC 10k thermistors in temperature of 100 C [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/hexp-sy19], and [Maśnicki, R., & Pałczyńska, B. (2021). Calculations of the resistance values of 10 ratio resistors [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/n324-jc97].

Illustration of the publication

Fig. 1. The measurement setup

 

Specifically, the values of material factors of individual thermistors were calculated using the results from the files Rt0.0C.lvm, RtII0.0C.lvm, Rt100.0C.lvm, RtII100.0C.lvm and resR.lvm.

The results are shown in the attached files: IB.lvm and IIB.lvm for two sets of 10 thermistors, respectively.

 

Illustration of the publication

                                                                                                              Fig. 2. The sensor conditioning unit

Dataset file

Material_factors.zip
1.1 kB, S3 ETag 40594dec5fb25baa02491bd62b1546c6-1, downloads: 77
The file hash is calculated from the formula
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} where a single part of the file is 512 MB in size.

Example script for calculation:
https://github.com/antespi/s3md5
download file Material_factors.zip

File details

License:
Creative Commons: by-nc 4.0 open in new tab
CC BY-NC
Non-commercial

Details

Year of publication:
2021
Verification date:
2021-05-31
Creation date:
2021
Dataset language:
English
Fields of science:
  • Automation, electronic and electrical engineering (Engineering and Technology)
DOI:
DOI ID 10.34808/exdv-6t87 open in new tab
Series:
Verified by:
Gdańsk University of Technology

Keywords

References

Cite as

seen 249 times