Analysis of exhaled breath for dengue disease detection by low-cost electronic nose system - Publication - Bridge of Knowledge

Search

Analysis of exhaled breath for dengue disease detection by low-cost electronic nose system

Abstract

This paper presents a procedure and a set-up of an electronic nose system analyzing exhaled breath to detect the patients suffering from dengue – a mosquito-borne tropical disease. Low-power resistive gas sensors (MiCS-6814, TGS8100) were used to detect volatile organic compounds (VOCs) in the exhaled breath. The end-tidal phase of patients exhaled breath was collected with a BioVOCTM breath sampler. Two strategies were assessed for breath samples measurement: either direct transfer from the BioVOCTM into the sensors test chamber, or storage in Tenax TA sorbent tubes followed by VOCs release through thermal desorption and then transfer into sensors test chamber. DC sensor resistances were recorded and processed by multivariate classifier algorithms to detect infected patients. The experimental studies were run on a group of 26 individuals (16 dengue diagnosed patients and 10 control volunteers). The detection accuracy of dengue patients was over 90%.

Citations

  • 1 0

    CrossRef

  • 0

    Web of Science

  • 1 3

    Scopus

Cite as

Full text

download paper
downloaded 14 times
Publication version
Accepted or Published Version
DOI:
Digital Object Identifier (open in new tab) 10.1016/j.measurement.2022.110733
License
Creative Commons: CC-BY-NC-ND open in new tab

Keywords

Details

Category:
Articles
Type:
artykuły w czasopismach
Published in:
MEASUREMENT no. 190,
ISSN: 0263-2241
Language:
English
Publication year:
2022
Bibliographic description:
Smulko J., Chludziński T., Majchrzak T., Kwiatkowski A., Borys S., Lisset Jaimes-Mogollón A., Manuel Durán-Acevedo C., Geovanny Perez-Ortiz O., Ionescu R.: Analysis of exhaled breath for dengue disease detection by low-cost electronic nose system// MEASUREMENT -Vol. 190, (2022), s.110733-
DOI:
Digital Object Identifier (open in new tab) 10.1016/j.measurement.2022.110733
Sources of funding:
  • COST_FREE
Verified by:
Gdańsk University of Technology

seen 138 times