Calibration of NTC 10k thermistors in temperature of 50 C - Open Research Data - MOST Wiedzy

Wyszukiwarka

Calibration of NTC 10k thermistors in temperature of 50 C

Opis

The presented data set is part of the research aimed at determining the actual characteristics of each thermistor in a package of twenty NTC10k type sensors.

The measurement system is shown in Fig. 1. A measurement system based on NI PXIe-1082Q (National Instruments Chassis), equipped with NI PXIe-8135 (Embedded Controller), NI PXIe-6341 (DAQ card) and NI SCB-68A (68-pin connector block), was used for the acquisition of measurement data. Eleven input channels of the DAQ card were used in the measurements. Channel 0 (ch_0) was used to measure the supply voltage. Sensors numbered 1 to 20 were tested in two parts, ten sensors per set. They were connected to the inputs of channels 1-10 (ch_1-ch_10) of the DAQ card.

Ilustracja publikacji

Fig. 1. The measurement setup

 

The conditioning unit for individual sensors is shown in Fig.2. These circuits are performed on connector block SCB-68A as the voltage dividers. For the set of ten thermistors, the supply voltage is measured in channel 0 (ch_0) of the DAQ card, and individual thermistors are connected to channels 1-10 (ch_i).

The thermistor set was placed in the T-704 thermocalibrator chamber at 50°C (confirmed by a digital temperature gauge DT-34 with an accuracy of 0.1°C).

Ilustracja publikacji

                                                                                                              Fig. 2. The sensor conditioning unit

The folder I50,0C  composed of 15 files 50,0Cx.lvm (where: x is from 0 to 14), contains the measurement results of supply voltage Uz (ch_0) (Fig.2) and voltages at the outputs of ten voltage dividers (ch_1 to ch_10), repeated in a series of 30 measurements for all 11 channels.

The measuring series was carried out also for the second package of ten thermistors. The folder was marked as II50,0C, containing the files II50,0Cx.lvm  (where: x is from 0 to 14).

Plik z danymi badawczymi

temp_50.zip
38.6 kB, S3 ETag c892dc1896efa4a0f9bf95f5c177268a-1, pobrań: 76
Hash pliku liczony jest ze wzoru
hexmd5(md5(part1)+md5(part2)+...)-{parts_count} gdzie pojedyncza część pliku jest wielkości 512 MB

Przykładowy skrypt do wyliczenia:
https://github.com/antespi/s3md5
pobierz plik temp_50.zip

Informacje szczegółowe o pliku

Licencja:
Creative Commons: by 4.0 otwiera się w nowej karcie
CC BY
Uznanie autorstwa
Dane surowe:
Dane zawarte w datasecie nie zostały w żaden sposób przetworzone.

Informacje szczegółowe

Rok publikacji:
2021
Data zatwierdzenia:
2021-06-30
Data wytworzenia:
2021
Język danych badawczych:
angielski
Dyscypliny:
  • Automatyka, elektronika i elektrotechnika (Dziedzina nauk inżynieryjno-technicznych)
DOI:
Identyfikator DOI 10.34808/mgtg-cc76 otwiera się w nowej karcie
Seria:
Weryfikacja:
Politechnika Gdańska

Słowa kluczowe

Powiązane zasoby

Cytuj jako

wyświetlono 141 razy