Optimization of a Fabry-Perot Sensing Interferometer design for an optical fiber sensor of hematocrit level - Publication - Bridge of Knowledge

Search

Optimization of a Fabry-Perot Sensing Interferometer design for an optical fiber sensor of hematocrit level

Abstract

Continuous measurement of the hemato crit level in blo o d can p otentially b e p erformed using optical bre sensors. The FabryPerot interferometric sensors are a promising candidate in this application. The most imp ortant step in the design of the sensor is design of the sensing interferometer. Adequate cavity length and high interference contrast are two most imp ortant requirements in this application. The design metho d of the sensing interferometer presented in this pap er uses a Gaussian b eam approximation. In order to verify the design, an optical bre Fabry Perot interferometer with adjustable cavity length was built. Its p erformance was tested, conrming validity of the design approach.

Citations

  • 1 7

    CrossRef

  • 0

    Web of Science

  • 2 1

    Scopus

Keywords

Details

Category:
Articles
Type:
artykuł w czasopiśmie wyróżnionym w JCR
Published in:
ACTA PHYSICA POLONICA A no. 124, pages 586 - 588,
ISSN: 0587-4246
Language:
English
Publication year:
2013
Bibliographic description:
Wierzba P., Jędrzejewska-Szczerska M.: Optimization of a Fabry-Perot Sensing Interferometer design for an optical fiber sensor of hematocrit level// ACTA PHYSICA POLONICA A. -Vol. 124, nr. 3 (2013), s.586-588
DOI:
Digital Object Identifier (open in new tab) 10.12693/aphyspola.124.586
Verified by:
Gdańsk University of Technology

seen 117 times

Recommended for you

Meta Tags