Abstract

Abdominal prostheses are used in humans for abdominal wall strengthening or reconstruction. From the mechanical point of view they are membranes, so their basic mechanical property is tensile stiffness. This property is experimentally identified for a selected implant, OptomeshTM. The obtained results are described herein. Uni-axial simple and cyclic tension tests are a basis for the analysis. Both kinds of tests allowed to distinguish two states of the material, namely: baseline and preconditioned. In the preconditioned state the material reveals approximate two-fold increase of tensile stiffness in comparison to baseline state. In both states the material demonstrate orthotropy. Both features are typical for knitted synthetic abdominal implants, however, the first one is not, while the other is desired from the mechanical point of view of operated hernia.

Citations

  • 0

    CrossRef

  • 0

    Web of Science

  • 0

    Scopus

Cite as

Full text

download paper
downloaded 42 times
Publication version
Accepted or Published Version
License
Copyright (2019 Author(s))

Keywords

Details

Category:
Articles
Type:
publikacja w in. zagranicznym czasopiśmie naukowym (tylko język obcy)
Published in:
AIP Conference Proceedings no. 2077, pages 020058 - 020065,
ISSN: 0094-243X
Language:
English
Publication year:
2019
Bibliographic description:
Tomaszewska A., Szinvai S., Żerdzicki K.. Stiffness change of abdominal prosthesis Optomesh™ under cyclic loading. AIP Conference Proceedings, 2019, Vol. 2077, iss. 1, s.020058-020065
DOI:
Digital Object Identifier (open in new tab) 10.1063/1.5091919