Opis
The presented data set is part of the research aimed at determining the actual characteristics of each thermistor in a package of twenty NTC10k type sensors.
The measurement system is shown in Fig. 1. A measurement system based on NI PXIe-1082Q (National Instruments Chassis), equipped with NI PXIe-8135 (Embedded Controller), NI PXIe-6341 (DAQ card) and NI SCB-68A (68-pin connector block), was used for the acquisition of measurement data. Eleven input channels of the DAQ card were used in the measurements. Channel 0 (ch_0) was used to measure the supply voltage. Sensors numbered 1 to 20 were tested in two parts, ten sensors per set. They were connected to the inputs of channels 1-10 (ch_1-ch_10) of the DAQ card.
The conditioning unit for individual sensors is shown in Fig.2. These circuits are performed on connector block SCB-68A as the voltage dividers. For the set of ten thermistors, the supply voltage is measured in channel 0 (ch_0) of the DAQ card, and individual thermistors are connected to channels 1-10 (ch_i).
The calculations of the material factors were performed based on the following datasets [Maśnicki, R., Tyndzik, K., & Pałczyńska, B. (2021). Calibration of NTC 10k thermistors in temperature of 0 C [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/dcn8-z120], [Maśnicki, R., Tyndzik, K., & Pałczyńska, B. (2021). Calibration of NTC 10k thermistors in temperature of 100 C [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/hexp-sy19], and [Maśnicki, R., & Pałczyńska, B. (2021). Calculations of the resistance values of 10 ratio resistors [Data set]. Gdańsk University of Technology. https://doi.org/10.34808/n324-jc97].
Fig. 1. The measurement setup
Specifically, the values of material factors of individual thermistors were calculated using the results from the files Rt0.0C.lvm, RtII0.0C.lvm, Rt100.0C.lvm, RtII100.0C.lvm and resR.lvm.
The results are shown in the attached files: IB.lvm and IIB.lvm for two sets of 10 thermistors, respectively.
Fig. 2. The sensor conditioning unit
Plik z danymi badawczymi
hexmd5(md5(part1)+md5(part2)+...)-{parts_count}
gdzie pojedyncza część pliku jest wielkości 512 MBPrzykładowy skrypt do wyliczenia:
https://github.com/antespi/s3md5
Informacje szczegółowe o pliku
- Licencja:
-
otwiera się w nowej karcieCC BY-NCUżycie niekomercyjne
Informacje szczegółowe
- Rok publikacji:
- 2021
- Data zatwierdzenia:
- 2021-05-31
- Data wytworzenia:
- 2021
- Język danych badawczych:
- angielski
- Dyscypliny:
-
- Automatyka, elektronika i elektrotechnika (Dziedzina nauk inżynieryjno-technicznych)
- DOI:
- Identyfikator DOI 10.34808/exdv-6t87 otwiera się w nowej karcie
- Seria:
- Weryfikacja:
- Politechnika Gdańska
Słowa kluczowe
Powiązane zasoby
- dane badawcze Calibration of NTC 10k thermistors in temperature of 0 C
- dane badawcze Calibration of NTC 10k thermistors in temperature of 100 C
- dane badawcze Calculations of the resistance values of 10 ratio resistors
- dane badawcze Determination of the ratio resistors in the input circuits of the NTC10k thermistor calibration system
- dane badawcze Calculations of the resistance values of 20 thermistors at 100°C
- dane badawcze Calculations of the resistance values of 20 thermistors at 0°C
Cytuj jako
Autorzy
wyświetlono 246 razy