Abstrakt
Abstract — A novel approach for monitoring a temperature distribution inside a tissue during thermoablation is presented in the paper. A thermal profile is measured using a set of serially connected thermistors each bypassed by a capacitor. This technique allows a two-wire and simultaneous multi- point measurements using a multi-frequency measurement of electrical impedance. It is shown that application signals of appropriately selected frequency allows simultaneous measure- ment of temperature at five distinct points. This technique can be utilized in the assisting of a thermoablation process, and in other applications based on resistive or capacitive sensors.
Cytowania
-
0
CrossRef
-
0
Web of Science
-
0
Scopus
Autorzy (2)
Cytuj jako
Pełna treść
- Wersja publikacji
- Accepted albo Published Version
- Licencja
- Copyright (2017 IEEE)
Słowa kluczowe
Informacje szczegółowe
- Kategoria:
- Aktywność konferencyjna
- Typ:
- publikacja w wydawnictwie zbiorowym recenzowanym (także w materiałach konferencyjnych)
- Tytuł wydania:
- 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) strony 4541 - 4544
- Język:
- angielski
- Rok wydania:
- 2017
- Opis bibliograficzny:
- Bujnowski A., Wtorek J.: A novel sensor for measuring temperature profile during the thermoablation// 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)/ ed. Jim Patton Seogwipo, South Korea : , 2017, s.4541-4544
- DOI:
- Cyfrowy identyfikator dokumentu elektronicznego (otwiera się w nowej karcie) 10.1109/embc.2017.8037866
- Weryfikacja:
- Politechnika Gdańska
wyświetlono 93 razy
Publikacje, które mogą cię zainteresować
Reliability of acicular grindable thermocouples for transient temperature measurements at sliding contacts
- O. Nosko,
- Y. Tsybrii,
- W. Tarasiuk
- + 1 autorów