Abstrakt

OBJECTIVES: To preclinically evaluate drug-eluting biopsy needles (patent pending WO2016118026) as a new potential way of antimicrobial prophylaxis for transrectal prostate biopsy. METHODS: Twenty steel biopsy needles have been coated with polyvinyl alcohol, ciprofloxacin, and amikacin. Modified biopsy needles have been randomly divided into 3 groups (1:2:1 ratio). Needles from group I were immersed for 30 minutes in dedicated test tubes containing saline. Needles from group II were immersed (one by one) for 5 seconds in a set of 12 test tubes containing saline. Then, each solution was analyzed using high-performance liquid chromatography. The results were compared with the susceptibility break points for Escherichia coli. Group III was incubated with E coli strains on Mueller-Hinton plate and then the bacterial inhibition zones surrounding needles were measured. RESULTS: The average concentration of antibiotics eluted from needles (group I) was 361.98 ± 15.36 µg/mL for amikacin and 63.87 ± 5.95 µg/mL for ciprofloxacin. The chromatographic analysis revealed the gradual release of both antibiotics from needles (group II). The concentration of amikacin released from needles exceeded the break-point value from first to ninth immersion. Ciprofloxacin concentration was higher than break-point value in all immersions. The average bacterial inhibition zone minor axis was 42 ± 5.7 mm (group III). CONCLUSIONS: The use of drug-eluting biopsy needle could be a new potential way of antimicrobial prophylaxis for transrectal prostate biopsy. This study confirmed its biological activity as well as the gradual release of antibiotics from its surface. Confirmation of its preventive role, in terms of infectious complications after transrectal prostate biopsy, has to be evaluated in a clinical trial.

Cytowania

  • 3

    CrossRef

  • 0

    Web of Science

  • 3

    Scopus

Autorzy (7)

Słowa kluczowe

Informacje szczegółowe

Kategoria:
Publikacja w czasopiśmie
Typ:
artykuł w czasopiśmie wyróżnionym w JCR
Opublikowano w:
TECHNOLOGY IN CANCER RESEARCH & TREATMENT nr 16, wydanie 6, strony 1038 - 1043,
ISSN: 1533-0346
Język:
angielski
Rok wydania:
2017
Opis bibliograficzny:
Sieczkowski M., Gibas A., Wasik A., Kot-Wasik A., Piechowicz L., Namieśnik J., Matuszewski M.: Drug-Eluting Biopsy Needle as a Novel Strategy for Antimicrobial Prophylaxis in Transrectal Prostate Biopsy// TECHNOLOGY IN CANCER RESEARCH & TREATMENT. -Vol. 16, iss. 6 (2017), s.1038-1043
DOI:
Cyfrowy identyfikator dokumentu elektronicznego (otwiera się w nowej karcie) 10.1177/1533034617722080
Weryfikacja:
Politechnika Gdańska

wyświetlono 181 razy

Publikacje, które mogą cię zainteresować

Meta Tagi