Application and optimization of headspace solid-phase microextraction (HS-SPME) coupled with gas chromatography - flame-ionization detector (GC-FID) to determine products of the petroleum industry in aqueous samples
Abstrakt
A simple and solvent-free analytical method has been optimized for the determination of hydrocarbons (C6 - C20) from products of the petroleum industry in aqueous samples. Extraction and analysis was carried out using headspace solid-phase microextraction (HS-SPME) and gas chromatography coupled with flame-ionization detector (GC-FID). A ZB-5 capillary column (30 m × 0.25 mm inner diameter, 0.25 µm film thickness) was used for GC separation and a 100 µm polydimethylsiloxane (PDMS) fiber was used to isolate the analytes. Parameters of HS-SPME, analytes desorption and GC-FID were selected and an analytical procedure is proposed. Under the optimum conditions the limits of detection were in the range of 2.0 - 13 µg L-1 and the relative standard deviations were between 2.6 - 14%. The developed analytical method was successfully applied to the analysis of water contaminated by unleaded petrol and unfiltered water used to extinguish a simulated fire.
Autorzy (3)
Cytuj jako
Pełna treść
pełna treść publikacji nie jest dostępna w portalu
Słowa kluczowe
Informacje szczegółowe
- Kategoria:
- Publikacja w czasopiśmie
- Typ:
- artykuł w czasopiśmie wyróżnionym w JCR
- Opublikowano w:
-
MICROCHEMICAL JOURNAL
nr 108,
strony 117 - 123,
ISSN: 0026-265X - Język:
- angielski
- Rok wydania:
- 2013
- Opis bibliograficzny:
- Tankiewicz M., Morrison C., Biziuk M.: Application and optimization of headspace solid-phase microextraction (HS-SPME) coupled with gas chromatography - flame-ionization detector (GC-FID) to determine products of the petroleum industry in aqueous samples// MICROCHEMICAL JOURNAL. -Vol. 108, (2013), s.117-123
- Weryfikacja:
- Politechnika Gdańska
wyświetlono 153 razy